NIR256 Temp-regulated NIR Spectrometers

NOTE: This item has been discontinued. Please click here for replacement information.

The NIR256-2.1 and NIR256-2.5 Near-infrared Spectrometers are designed for applications that require sensitivity in the near-infrared region such as tunable laser wavelength characterization and general NIR spectroscopy.

Available for  $23099, the NIR256-2.1 Spectrometer features a temperature-regulated 256-element InGaAs array detector and is effective in the 0.9-2.1µm wavelength range or 1.2-2.1 µm range, depending on the grating selected.

The NIR256-2.5 extends farther into the NIR, acquiring spectra up to 2.5 µm. The NIR256-2.5 is available for $25405.

Applications

  • Sugar analysis

  • Moisture analysis

  • Pulpwood QC

  • Alcohol analysis in brewing

  • Nitrogen detection in soil

  • Fat, oil and lipid determination

NIR256-2.1 Spectrometer: Spectra to 2.1 µm  

The NIR256 uses a 256-element InGaAs linear-array detector. With the NIR256 you have two grating options. With Grating N1, you have a 1200-2100 nm wavelength range. Grating N2 provides a 900-2100 nm wavelength range. The NR256-2.1 acquires data as fast as 10 milliseconds.

NIR256-2.5 Spectrometer: Spectra to 2.5 µm

The NIR256-2.5 Extended-range Spectrometer extends farther into the NIR, acquiring spectra up to 2.5 um. The NIR256-2.5 features a temperature-regulated InGaAs detector array, which is internally cooled for optimum signal-to-noise and sensitivity. The NR256-2.5 acquires data as fast as 10 milliseconds.

"Cool Design"

Each InGaAs detector has onboard thermoelectric cooling. A thermistor monitors the array's temperature and a thermoelectric device can cool the array to 30 °C below ambient, keeping the array stable to within +/-0.1 °C. You can set and monitor the detector's temperature via software.

Integrated Design

A 16-bit A/D converter is mounted with the spectrometer in the same housing. A 5 VDC wall transformer (included) is required. When an NIR256 Spectrometer inerfaces to a PC via the USB port, you have access to the spectrometer's EEPROM, where wavelength calibration coefficients and other information unique to your spectrometer is stored. Interface via an RS-232 serial port interface is also possible.

Convenient Calibration

To maintain the quality of your spectrometer's wavelength calibration, we recommend our drift-free, light sources specialized for spectrometer wavelength calibration. Click here for details.

Specifications

Specification NIR 256-2.1 NIR 256-2.5
Physical
Dimensions: 153.4 mm x 105.2 mm x 76.2 mm 153.4 mm x 105.2 mm x 76.2 mm
Weight: 980 grams 980 grams
Detector

Detector:

Hamamatsu G9206-256; InGaAs linear array

Hamamatsu G9208-256; InGaAs linear array

Detector range: 900-2100 nm 900-2550 nm

Pixels:

256 pixels*

256 pixels*

Pixel size:

50 μm x 250 μm

25 μm x 250 μm

Pixel well depth:

188 Me- electrons

188 Me- electrons

Defective pixels: 2% 5%
Optical Bench
Design: f/4, Symmetrical crossed Czerny-Turner f/4, Symmetrical crossed Czerny-Turner
Focal length: 40 mm input 40 mm input
Entrance aperture: 10, 25, 50, 100 or 200 µm wide slits or fiber (no slit) 10, 25, 50, 100 or 200 µm wide slits or fiber (no slit)
Grating options: 2 grating options (N1 or N2 Grating) 1 grating option (N2 Grating)
HC-1 grating option: No No
Detector collection lens option: No No
OFLV filter options: No No
Other bench filter options: No No
Collimating and focusing mirrors: Standard Standard
UV enhanced window: No No
Fiber optic connector: SMA 905 to 0.22 numerical aperture single-strand optical fiber SMA 905 to 0.22 numerical aperture single-strand optical fiber
Spectroscopic
Wavelength range: 900-2100 nm with Grating N2; 1200-2100 nm with Grating N1 900-2500 nm with Grating N2
Optical resolution: ~12.0 nm FWHM ~7.5-25.0 nm FWHM
Signal-to-noise ratio: 4000:1 (at full signal) 4000:1 (at full signal)
A/D resolution: 16 bit 16 bit
Dark noise: 12 RMS counts 12 RMS counts
Dynamic range: 5 x 10^6 (system); 5000:1 for a single acquisition 5 x 10^6 (system); 5000:1 for a single acquisition
Integration time: 1 millisecond to 1 second 1 millisecond to 30 milliseconds
Corrected linearity: >99.6% >99.6%
Electronics
Power consumption: 3 A @ 5 VDC 2 A @ 5 VDC
Data transfer speed: Full scans to memory every 10 ms with USB port Full scans to memory every 10 ms with USB port
Inputs/Outputs: No No
Analog channels: No No
Auto nulling: No No
Breakout box compatibility: No No
Trigger modes: 3 modes 3 modes
Strobe functions: Yes Yes
Gated delay feature: No No
Connector: 15-pin connector 15-pin connector
Computer
Operating systems: Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port
Computer interfaces: USB 2.0 @ 12 Mbps; RS-232 (2-wire) @ 115.2 K baud USB 2.0 @ 12 Mbps; RS-232 (2-wire) @ 115.2 K baud
Peripheral interfaces: I2C inter-integrated circuit; SPI (3-wire) I2C inter-integrated circuit; SPI (3-wire)
Pricing
Retail $23099 $25405

Software Interface

Use the NIR256 with SpectraSuite Spectroscopy Platform Software, which is available for $199.

* Download free software to analyze your computer's USB readiness by clicking here.

.

HOME PRODUCTS SERVICES TECHNICAL APPLICATIONS CORPORATE DISTRIBUTORS CATALOG NEWSLETTER CONTACT

.
Copyright 1989 - 2010 Ocean Optics, Inc. All rights reserved.  Terms of Use and Privacy Statement
+1 727.733.2447 • +1 Fax 727.733.3962 •
 

 

Contact an Ocean Optics Applications Scientist

 

Convenient Wavelength Calibration! Click here for Details

 

NIR512 Spectrometer

LS-1 Tungsten Halogen Light Source

NIR256 Annual Service Package