Physical |
|
| Dimensions: | 153.4 mm x 105.2 mm x 76.2 mm |
| Weight: | 980 grams |
| Detector | |
|
Detector: |
Hamamatsu G9204-512; InGaAs linear array |
| Detector range: | 850-1700 nm |
|
Pixels: |
512 pixels |
|
Pixel size: |
25 μm x 500 μm |
|
Pixel well depth: |
188 Me- electrons |
| Optical Bench | |
| Design: | f/4, Symmetrical crossed Czerny-Turner |
| Focal length: | 40 mm input |
| Entrance aperture: | 5, 10, 25, 50, 100 or 200 µm wide slits or fiber (no slit) |
| Grating options: | 1 grating option (N1 Grating) |
| HC-1 grating option: | No |
| Detector collection lens option: | No |
| OFLV filter options: | No |
| Other bench filter options: | No |
| Collimating and focusing mirrors: | Standard |
| UV enhanced window: | No |
| Fiber optic connector: | SMA 905 to 0.22 numerical aperture single-strand optical fiber |
| Spectroscopic | |
| Wavelength range: | 900-1700 nm with N1 Grating |
| Optical resolution: | ~3.0 nm FWHM |
| Signal-to-noise ratio: | 4000:1 (at full signal) |
| A/D resolution: | 16 bit |
| Dark noise: | 12 RMS counts |
| Dynamic range: | 5 x 10^6 (system); 5000:1 for a single acquisition |
| Integration time: | 1 millisecond to 3 seconds |
| Corrected linearity: | >99.8% |
| Electronics | |
| Power consumption: | 2 A @ 5 VDC |
| Data transfer speed: | Full scans to memory every 10 ms with USB port |
| Inputs/Outputs: | No |
| Analog channels: | No |
| Auto nulling: | No |
| Breakout box compatibility: | No |
| Trigger modes: | 3 modes |
| Strobe functions: | Yes |
| Gated delay feature: | No |
| Connector: | 15-pin connector |
| Computer | |
| Operating systems: | Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
| Computer interfaces: | USB 2.0 @ 12 Mbps; RS-232 (2-wire) @ 115.2 K baud |
| Peripheral interfaces: | I2C inter-integrated circuit; SPI (3-wire) |
| Pricing | |
| Retail | $17320 |
NIR-512 Grating Efficiency Curve
The NIR-512 has a grating with a groove density of 300 mm-1 blazed at 1100 nm.
|
|
Use the NIR512 with SpectraSuite Operating Software, which is available for $199.
* Download free software to analyze your computer's USB readiness by clicking here.
PRODUCTS |
TECHNICAL |
APPLICATIONS |
CORPORATE |
DISTRIBUTORS
| PRIVACY POLICY
CONTACT US |
CATALOG |
PRICE LIST |
ORDER INFO

.
©2007 Ocean Optics, Inc. All rights reserved.
Terms of
Use and Privacy Statement
+1
727.733.2447 • +1 Fax 727.733.3962 •
![]()
Contact an
Ocean Optics
Applications Scientist
Convenient Wavelength Calibration! Click here for Details
LS-1 Tungsten Halogen Light Source