Ocean Optics - Inventor of the World's First Miniature Spectrometer
Ocean Optics - Inventor of the World's First Miniature Spectrometer
 

Ocean Optics - Inventor of the World's First Miniature Spectrometer

 

SF2000 Spectrofluorometer

Conceived for use with our Fiber Optic Oxygen Sensors, the SF2000 Miniature Fiber Optic Spectrophotometer is a low-cost, high-sensitivity spectrometer configured especially for measuring fluorescence.

The SF2000 is no longer available. It has been replaced with several other fluorescence spectrometers:

At the heart of the SF2000 is an S2000 Series Fiber Optic Spectrometer pre-configured to a 360-1000 nm wavelength range (click here for grating efficiency), with a 200 µm entrance slit (for optical resolution to ~10.0 nm FWHM ) and L2 Detector Collection Lens (for increased light throughput). Unlike the S2000-FL Spectrometer for fluorescence, which is similarly configured, the SF2000 also features an integral R-LS-450 card-mounted Pulsed Blue LED. Both spectrometer and LED are mounted in an SD2000 dual-channel spectrometer housing (see photo).

The SF2000 interfaces to a desktop PC via Ocean Optics' ADC1000 ISA-bus A/D converter. Another interface option is the ADC1000-USB, an external A/D converter with 1 MHz sampling frequency that connects to the spectrometer via a DB25-pin connector and to desktop or notebook PCs via the USB port. A PCI-bus converter is also available.

The SF2000's integral R-LS-450 Pulsed Blue LED uses an external power source but has internal connections for synchronization with the spectrometer.

The SF2000 Spectrometer has a compact optical bench coupled to a high-sensitivity 2048-element linear CCD-array detector. The spectrometer accepts light energy transmitted through single-strand optical fiber and disperses it via a fixed grating (set for 360-1000 nm) across the CCD array. An SMA 905 connector allows for easy coupling to Fiber Optic Oxygen Sensors and other fiber optic accessories.

Specifications

Detector Specifications

Detector: 2048-element linear silicon CCD array
CCD elements: 2048 elements @ 14 µm x 200 µm per element
Well depth (@ 600 nm): 160,000 photons
Sensitivity (estimated):
  • 86 photons/count
  • 2.9 x 10-17 joule/count
  • 2.9 x 10-17 watts/count (for 1-second integration)
Detector range: 200-1100 nm
Useable range: 360-1000 nm
Integration time: 4 milliseconds to 60 seconds (with 500 kHz A/D converter)
20 milliseconds to 60 seconds (with 100 kHz PCMCIA A/D converter)

For the SF2000, performance specifications reflect a configuration that includes a 600 lines/mm grating, 200 µm entrance slit and L2 Detector Collection Lens.

Optics Specifications

Grating: 600 lines/mm, set to 360-1000 nm (blaze wavelength is 500 nm)
Slit: 200 µm width (height is 1 mm)
Resolution: ~10.0 nm (FWHM)
Stray light: < 0.05% at 600 nm
< 0.10% at 435 nm
Fiber optic connector: SMA 905 to single-strand optical fiber (0.22 NA)

.

HOME PRODUCTS SERVICES TECHNICAL APPLICATIONS CORPORATE DISTRIBUTORS CATALOG NEWSLETTER CONTACT

.
Copyright 1989 - 2010 Ocean Optics, Inc. All rights reserved.  Terms of Use and Privacy Statement
+1 727.733.2447 • +1 Fax 727.733.3962 •

 

 


Contact an Ocean Optics Applications Scientist

 

Convenient Wavelength Calibration!
Click here for Details