USB2000-UV-VIS Spectrometer
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Cable |
Description |
Price |
|
USB-CBL-1 |
Cable connects from USB port on USB2000 to USB port on desktop or notebook PC |
$25 |
|
USB-ADP-PC |
Cable and adapter block connect from serial port on USB2000 to serial port on desktop or notebook PC; also comes with USB-CBL-PS 5 VDC power supply |
$75 |
|
CBL-SER |
Spare 9-pin serial cable; connects to USB-ADP-H and serial port on desktop or notebook PC |
$25 |
|
USB-CBL-PS |
Spare 5 VDD power supply; required if USB2000 interfaces to PC via serial port |
$25 |
|
USB-ADP-H |
Adapter block to use with CBL-SER to connect from serial port on USB2000 to serial port on desktop or notebook PC |
$75 |
|
USB-ADP-PX2 |
Adapter block and cable to conect the PX-2 Pulsed Xenon Lamp or LS-450 Pulsed Blue LED to the USB2000 |
$50 |
A pair of Lithium Ion cells makes the USB2000-UV-VIS Spectrometer field-portable. The USB-BP Battery Pack delivers 225 mA at 5 Volts -- enough juice to power a light source and the USB2000 for 8 hours continuously, or a USB-LS-450 LED module and the USB2000 for 12 hours continuously. Included is a battery charger for replenishing spent batteries.
We've developed a multi-purpose LED drive module for USB2000 Spectrometers that has a built-in 24-bit A/D converter to accommodate a 100 ohm Resistance Temperature Device for temperature measurements (temperature probe priced separately). Conceived for use with Fiber Optic Oxygen Sensors, the USB-LS-450 LED Light Source Module also features on-board memory to store temperature and oxygen calibration coefficients and is available with an optional add-on that provides 4-20 mA output for process and other applications.
|
Computer interface: |
Universal Serial Bus (RS-232 available on side connector) |
|
Spectrometer channels: |
Master spectrometer channel only |
|
Computer requirements*: |
|
|
Integration time: |
3 milliseconds-65 seconds |
|
Data transfer rate: |
|
|
Dimensions: |
3.5" x
2.5"x 1.31" LWH |
|
Weight: |
0.45 lb.
without cable, 0.60 lb. with cable |
* Download free software to analyze your computer's USB readiness by clicking here.
|
Detector: |
2048-element linear silicon CCD array |
|
Effective range: |
200-1100 nm |
|
Dynamic range: |
2 x 108 |
|
Sensitivity (estimate): |
|
|
Signal-to-Noise: |
250:1 (at full signal) |
|
Dark noise: |
2.5-4.0 (RMS) |
|
Grating: |
600 lines/mm, set to 200-850 nm (blazed at 300 nm) |
|
Slit: |
25 mm width (height is 1000 mm) |
| Focal length: | 42 mm (input); 68 mm (output) |
| Order-sorting: | Single-piece, multi-bandpass detector coating to eliminate second-order effects from 200-850 nm |
| Resolution: | 1.5 nm (FWHM) |
| Stray light: | < 0.05% at 600 nm < 0.10% at 435 nm < 0.10% at 250 nm |
| Fiber optic connector: | SMA 905 to single-strand optical fiber (0.22 NA) |
| . | |||||||||
|
|
|||||||||
| HOME | PRODUCTS | SERVICES | TECHNICAL | APPLICATIONS | CORPORATE | DISTRIBUTORS | CATALOG | NEWSLETTER | CONTACT |
![]() ![]() ![]() . Copyright 1989 - 2010 Ocean Optics, Inc. All rights reserved. Terms of Use and Privacy Statement +1 727.733.2447 • +1 Fax 727.733.3962 • |
|||||||||
![]()
Contact an Ocean Optics Applications Scientist
Convenient Wavelength Calibration!