Predicted Ranges & Resolution: QE Bench

These graphs demonstrate the range and resolution of your "QE" Bench Spectrometer with a 5 µm Slit and various gratings.

Example: If the starting wavelength is 400 nm, then the range is ~364 nm, providing a 400-764 nm wavelength range and 0.645 nm optical resolution (FWHM). Gratings for the QE Bench that have groove densities of 600 mm-1 are Grating H1, H2, H3, H4 and H14.

 

Example: If the starting wavelength is 600 nm, then the range is ~152 nm, providing a 600-752 nm wavelength range and 0.295 nm optical resolution (FWHM). Gratings for the QE Bench that have groove densities of 1200 mm-1 are Grating H5, H6 and H9.

 

Example: If the starting wavelength is 300 nm, then the range is ~112 nm, providing a 300-412 nm wavelength range and 0.226 nm optical resolution (FWHM). Gratings for the QE Bench that have groove densities of 1800 mm-1 are Grating H10 and H11.

 

Example: If the starting wavelength is 450 nm, then the range is ~62 nm, providing a 450-512 nm wavelength range and 0.139 nm optical resolution (FWHM). Gratings for the QE Bench that have groove densities of 2400 mm-1 are Grating H7 and H12.

 

 

.