Probe Specifications: |
|
|
FOXY
Formulation |
FOSPOR Formulation |
HIOXY
Formulation |
| |
(general-purpose
coating) |
(high-sensitivity for
low levels of O2) |
(for
hydrocarbon
environments) |
|
O2%
range (at 1 ATM): |
0-100% |
0-10% |
0-20.9% |
|
DO
range (ppm at 1 ATM): |
0-40 ppm |
0-4 ppm |
0-8 ppm |
|
Temperature range: |
-50-+80 ºC for probes |
-50-+80 ºC for probes |
-50-+60
ºC |
|
0-60
ºC for patches |
0-60
ºC for patches |
NA |
|
O2%
resolution: |
0.01-0.05% |
.001% |
0.01-0.05% |
|
DO
resolution (at room
temp): |
4-20
ppb |
0.4-2
ppb |
4-20
ppb |
|
O2%
accuracy: |
5% of
reading |
5% of
reading |
5% of
reading |
|
DO
accuracy: |
5% of
reading |
5% of
reading |
5% of
reading |
|
Min.
detectable level in gas: |
0.01% - 0.05% |
0.001% - 0.01% |
0.01% - 0.05% |
|
Response time: |
< 1
second in gas |
< 1
second in gas |
< 1
second in gas |
|
~30
seconds with overcoating
in gas |
~30
seconds with overcoating
in gas |
~45
seconds in pure water |
|
|
~45
seconds in pure water |
~45
seconds in pure water |
|
|
Patch Specifications: |
|
|
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
|
02% range (at 1 ATM): |
0-100% |
0-20.9% |
0-100% |
|
DO range (ppm at 1 ATM): |
0-40 ppm |
0-8 ppm |
0-40 ppm |
|
Temperature range: |
-20 to +60 ºC for
patches |
-20 to +60 ºC for
patches |
-20 to +60 ºC for
patches |
|
02% resolution (20 sex
averaging): |
0.05% |
0.01% |
0.05% |
|
DO resolution (at room
temp): |
20 ppb |
4 ppb |
20 ppb |
|
02% accuracy: |
5% of reading |
5% of reading |
5% of reading |
|
DO accuracy: |
5% of reading |
5% of reading |
5% of reading |
|
Min. detectable level: |
0.1% O2 |
0.01% O2 |
0.1% O2 |
|
Min detectable level in
water (at room temp): |
40 ppb |
4 ppb |
40 ppb |
|
Response time: |
<1 second in gas |
<1 second in gas |
<1 second in gas |
|
|
~30 seconds with
overcoating in gas |
~30 seconds with
overcoating in gas |
|
|
|
~45 seconds in pure
water |
~45 seconds in pure
water |
~45 seconds in pure
water |
|
Sterilization Guide: |
|
|
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
|
Form factor: |
Probes/Patches |
Probes/Patches |
Probes |
|
Autoclavable: |
No |
No |
Yes |
|
Calibration: |
|
|
Single point reset
required after
sterilization |
|
Misc: |
|
|
|