S2000 Miniature Fiber Optic Spectrometer
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Product | Description | Power Consumption | A/D Converter | Price |
| S2000 (master) | Master spectrometer channel with housing | 110 mA @ 5 VDC | Additional | $1890 |
| S2000 (additional) | Additional spectrometer channel, no housing** | 60 mA @ 5 VDC | Additional | $838 |
| SD2000 | Two spectrometer channels in dual housing | 170 mA @ 5 VDC | Additional | $2599 |
| PC2000 | Master spectrometer channel on A/D converter to insert in PC | 250 mA @ 5 VDC | 1 MHz ISA-bus* | $1999 |
| PC2000-S-ISA | Additional spectrometer channel on A/D converter to insert in PC | 70 mA @ 5 VDC | 1 MHz ISA-bus* | $799 |
| PC2000-S-PCI | Additional spectrometer channel on A/D converter to insert in PC | 70 mA @ 5 VDC | 1 MHz PCI-bus | $799 |
| PC2000-PC/104 | Master spectrometer mounted onto a PC/104 A/D converter, for OEM customers only | 250 mA @ 5 VDC | 1 MHz PC/104 | $2199 |
In addition, the following upgrades are available:
| Item | Description | Price |
| UPGD-S-1 | S2000 series upgrade, one optical bench component | $400 |
| UPGD-S-2 | S2000 series upgrade, two optical bench components | $550 |
| UPGD-S-3 | S2000 series upgrade, three optical bench components | $700 |
| UPGD-S-4 | S2000 series upgrade, three optical bench components | $850 |
The S2000 performance specifications described here were achieved under the following conditions: The configuration featured a 600 lines/mm grating (set for 300-950 nm), with a tungsten-halogen light source for illumination and a 1000 µm optical fiber for signal collection. The configuration did not include an entrance slit. Integration times varied.
|
Dimensions: |
141.6 mm x 104.9 mm x 40.9 mm (with enclosure, master channel only) |
|
Weight: |
390 g (with enclosure, master channel only) |
|
Power consumption: |
110 mA @ 5 VDC (master); 60 mA @ 5 VDC (slave) |
|
Detector range: |
200-1100 nm |
|
Detector: |
2048-element linear silicon CCD array |
|
Gratings: |
14 gratings: UV through Shortwave NIR |
|
Entrance aperture: |
5, 10, 25, 50, 100 or 200 µm wide slits |
|
Order sorting filters: |
Installed bandpass and longpass filters |
|
Focal length: |
42 mm (input); 68 mm (output) |
|
Optical resolution: |
~0.3-10.0 nm FWHM |
|
Stray light: |
<0.05% at 600 nm; <0.10% at 435 nm; <0.10% at 250 nm |
|
Dynamic range: |
2 x 10 8 (system); 2000:1 for a single scan |
|
Sensitivity (estimated): |
86 photons/count; 2.9 x 10-17 joules/count; 2.9 x 10-17 watts/count (for 1-second integration) |
|
Fiber optic connector: |
SMA 905 to single-strand optical fiber (0.22 NA) |
|
Integration time: |
3 milliseconds to 60 seconds with 1
MHz A/D card |
PRODUCTS |
TECHNICAL |
APPLICATIONS |
CORPORATE |
DISTRIBUTORS
| PRIVACY POLICY
CONTACT US |
CATALOG |
PRICE LIST |
ORDER INFO

.
©2007 Ocean Optics, Inc. All rights reserved.
Terms of
Use and Privacy Statement
727.733.2447 • Fax 727.733.3962 •
![]()