SR6 Series Spectrometers
High Sensitivity and Great UV Response
High-sensitivity SR6 spectrometers offer excellent SNR and UV response for applications including UV absorbance, fluorescence and plasma monitoring. SR6 is a valued as a one-off setup for the lab or a customized system for high-volume industrial and OEM applications.
Documents
Manual
Drawing
Step File
Comms Protocol
SR6 Series
Disclaimer: Wavelength range and optical resolution specifications are typical values. Actual performance may vary depending on configuration, selected components, and operating conditions.
Group
Part Number
Wavelength Range
Optical resolution (FWHM in nm)
Grating Groove Density (lines/mm)
Grating Blaze Wavelength
Available at Edmund Optics25 μm slit only
5 μm slit 10 μm slit 25 μm slit 50 μm slit 100 μm slit 200 μm slit
UV-Vis options: SR-6UVV300-XX 180 – 900 nm 1.17 1.25 1.64 2.54 4.68 9.36 600 300 nm Buy Now
Vis-NIR options: SR-6VN500-XX 350 – 1040 nm 1.17 1.25 1.64 2.54 4.68 9.36 600 500 nm Buy Now
UV options: SR-6UV220-XX 180 – 400 nm 0.36 0.38 0.50 0.78 1.44 2.88 1800 Holographic —
SR-6UV240-XX 180 – 510 nm 0.54 0.58 0.76 1.17 2.16 4.32 1200 Holographic —
UV-Vis options: SR-6UVV240-XX 180 – 850 nm 1.17 1.25 1.64 2.54 4.68 9.36 600 240 nm —
SR-6UVV400-XX 180 – 900 nm 1.17 1.25 1.64 2.54 4.68 9.36 600 400 nm —
Vis options: SR-6VIS400-XX 350 – 1000 nm 1.17 1.25 1.64 2.54 4.68 9.36 600 400 nm —
Vis-NIR Ethernet options: SR-6VN500EN-XX 350 – 1040 nm — 1.25 1.64 — — — 600 500 nm —
NIR options: SR-6N750-XX 570 – 860 nm 0.54 0.58 0.76 1.17 2.16 4.32 1200 750 nm —
SR-6N1000-XX 600 – 1100 nm 1.17 1.25 1.64 2.54 4.68 9.36 600 1000 nm —
NIR Ethernet options: SR-6N750EN-XX 570 – 860 nm 0.54 0.58 — — — — 1200 750 nm —
Extended-range options: SR-6XR250-XX 190 – 1050 nm 1.44 1.54 2.02 3.12 5.76 11.52 500 250 nm —
Extended-range Ethernet options: SR-6XR250EN-XX 190 – 1050 nm — — 2.02 — — — 500 250 nm —
Integration time
7.2 ms – 5 s
SNR (single scan @ 10 ms)
400:1
SNR (max. per second w/High Speed Averaging Mode)
3,500:1
Connectors
16 pin Samtec TFM, RS-232, SMA, USB Type-C
Physical dimensions
88.1 x 63.5 x 31.45 mm
Weight
254 g
Temperature (operation)
0 °C – 55 °C
Temperature (storage)
-30 °C – 70 °C
Best for
UV absorbance of proteins
Fluorescence of chemical samples
High-sensitivity analysis