PX-2
Product Details
Xenon light sources offer very high intensity in the UV and are useful for absorbance, fluorescence or reflectance measurements.
The PX-2 pulsed xenon source has high intensity with a lower duty cycle, making it ideal for measurements where high intensity UV light can damage the sample. It has high flash rate, short-arc xenon lamp with output from 220-750 nm.
Product Overview
Pulse-to-Pulse Stability
Excellent performance for long-duration experiments
Adjustable Flash Rate
Optimize measurements of light sensitive samples
Simple Experiment Setup
Control of flash rate and synchronization via software
Specifications
Wavelength Coverage
220-750 nm
| Specifications | PX-2 Pulsed Xenon Lamp |
| Wavelength Range | 220-750 nm |
| Source | Pulsed Xenon |
| Nominal Bulb Power | NA |
| Typical Output Power | 9.9 W (average) |
| Source Lifetime | 10⁹ pulses (Estimated 230 days continuous operation @ 50 Hz pulse rate) |
| Warm-up Time | NA |
| Operating Humidity | 5-95% without condensation at 40 °C |
| Operating Temperature | 5 °C – 35 °C |
| Shutter | Yes |
| Fiber Connector | SMA 905 |
| Integrated Filter Holder | No |
| Dimensions | 14 cm x 10.5 cm x 4 cm |
| Weight | 370 g |
| Power Consumption | 1.3 A @ 11V @ 220 Hz 100 mA @ 12V @ 10Hz |
| Power Requirements | 85-264 V 50/60 Hz |
| Trigger/Shutter Input Connection | SUB-D-15 pin |
| Trigger/Shutter Input Signal | TTL 1-220 Hz |